If the loop requests a running TBR to be cancelled the Combo will set a TBR of 90% or 110% for 15 minutes instead.Similarly, when reducing to 50%, the lowest basal rate must be at least 0.10 U/h. when increasing to 200%, the highest basal rate must not exceed 5 U/h since it will be doubled. This also applies when modifying a profile, e.g. Currently only basal rates in the range of 0.05 to 10 U/h are supported.
#Download accu chek 360 software update
It’s currently not possible to set the time and date on the pump, so daylight saving time changes have to be performed manually (you may disable the phone’s automatic clock update in the evening and change it back in the morning together with the pump clock to avoid an alarm during the night).Setting a basal profile other than 1 on the pump or delivering extended boluses or multiwave boluses from the pump interferes with TBRs and forces the loop into low-suspend only mode for 6 hours as the the loop can’t run safely under these conditions.Extended bolus and multiwave bolus are not supported (see Extended Carbs instead).
#Download accu chek 360 software android
This allows using phones with Android < 8.1 but has not been widely tested:
#Download accu chek 360 software software
(Roche sends out Smartpix devices and the configuration software free of charge to their customers upon request.)